Bacharach AGM300 Smoke Alarm User Manual


 
60 Instruction 3015-4275
LINE 7: PPM – Parts Per Million is the volume concentration referenced to standard temperature and
pressure and is computed from lines 5, 8 and 13. There are two figures given. The first (annotated by a B)
is the actual PPM at the IR bench. The second is a PPM reading normalized to standard temperature and
pressure.
LINE 8: Zero Temperature – This line contains the sensor and enclosure temperature measured and
stored during the last purge cycle in degrees C.
LINE 9: Bench Temperature – This is the current IR sensor temperature in degrees C as well as the raw
voltage coming from the temperature sensor itself. This value can range from ambient to ambient +15
degrees C.
LINE 10: Box Temperature – This is the current internal enclosure temperature along with the raw
voltage from the temperature sensor, and has the same range as line 9.
LINE 11: Pressure – This is the current absolute manifold pressure in PSIA along with the output voltage
of the pressure sensor. This value should always be 0.2 to 1.0 PSIA below ambient (line 13).
LINE 12: Vacuum – The vacuum pressure is measured every purge cycle by closing all sample valves.
Its value is typically 2.5 to 4.0 PSIA below ambient (line 13).
LINE 13: Ambient – Ambient pressure is measured every purge cycle with the sample pump off and the
manifold open. Its value is weather and altitude dependent and can range from 10.0 to 15.5 PSIA.