Teledyne M201E Pet Fence User Manual


 
Maintenance Schedule Teledyne API Model M201E NH
3
Analyzer Operator Manual
38 05206H DCN5910
2. Remove the cover from the Converter chassis.
3. Unscrew the (4) nuts that secure the front panel to the chassis. They are located just
behind the Front Panel along the top.
4. Lower the Front Panel to gain easier access to the end of the quartz tube.
5. Unscrew the (3) nuts that secure the inner cover protecting the Heater Block and
quartz tube. Remove this cover.
6. Cut the tie-wrap that secures the thermocouple to the fitting at the end of the quartz
tube.
7. Loosen the Teflon fitting at the end of the quartz tube taking care not to put any
stress on the tube, and slide the fitting off the tube.
8. Remove the thermocouple.
9. Disconnect the thermocouple wires from the Temperature Controller noting that the
‘Yellow’ wire goes to pin 1 and the ‘Red’ wire goes to pin 2.
10. The thermocouple should slide into the Heater Block.
11. Replace the Tie-Wrap Hold-Down.
12. Form the thermocouple wire so that it rests in the cavity with little movement.
13. Connect the (2) wires of the thermocouple to the Temperature Controller. The
“yellow” wire should be connected to pin 1 and the “red” wire should be connected to
pin 2.
14. At this point, all connections have been made, both electrically and pneumatically. A
leak check should be performed on the Converter to verify that all connections are
leak free. If a leak is detected, the leak should be resolved before continuing.
15. Install the inner cover of the Heater Block and secure with the (3) nuts. Close the
Front Panel and secure with the (4) nuts. Install the top cover on the Converter
chassis.
16. The Converter is now ready for the application of power. You will be looking for an
indication from the Temperature Controller that it is functioning correctly and driving
the heater to the desired “set” temperature. Apply power now.
17. After the Converter comes to the regulated temperature, perform the Auto Tune
function to tune the Temperature Controller to the new thermocouple.
18. After the Auto-Tune process is completed, verify that the “process” temperature is
indicating that the desired temperature is stable and being regulated.
The converter is now ready for operation.