Teledyne M201E Pet Fence User Manual


 
Teledyne API Model M201E NH
3
Analyzer Operator Manual Configurable Analog Output Overview
05206H DCN5910 33
To make these adjustments the auto calibration feature “DATA_OUT_1 AUTO
CAL: OFF” must be turned off for each output channel then press the following:
DIAG AIO DATA_OUT_1 CALIBRATED: YES
< SET CAL EXIT
DIA
G
A
NALOG I / O CONFIGURATION
PREV NEXT ENTR EXIT
DIAG AIO
A
OUTS CALIBRATED: NO
< SET SET> CAL EXIT
FROM ANALOG I/O CONFIGURATION MENU
DIAG AIO DATA_OUT_1 RANGE: 5V
SET> EDIT EXIT
DIAG AIO DATA_OUT_1 CALIBRATED: N
O
< SET CAL EXIT
DIAG AIO DATA_OUT_1 VOLT
Z : 0 mV
U100 UP10 UP DOWN DN10 D100 ENTR EXIT
DIAG AIO DATA_OUT_1 :5V, TNXCNC1, NO CAL
< SET SET> EDIT EXIT
These keys
increment/decrement the
ZERO/SPAN D-to-A converter
output by
100, 10 or 1 counts
respectively.
Continue adjustments until the
voltage measured at the
output of the analyzer and/or
the input of the recording
device matches the value in
the upper right hand corner of
the display to the tolerance
listed previously.
The analyzer display WILL
NOT CHANGE. Only the
voltage reading of your volt
meter will chan
g
e.
DIAG AIO DATA_OUT_1 VOLT
S : 4500 mV
U100 UP10 UP DOWN DN10 D100 ENTR EXIT
EXIT ignores the
new setting.
ENTR accepts
the
new setting.
Press SET> to select the Analog Output
channel to be configured:
DISPLAYED AS = CHANNEL
DATA_OUT_1 = A1
DATA_OUT_2 = A2
DATA_OUT_3 = A3
DATA_OUT_4 = A4
Then Press EDIT to continue
Press SET Until
Figure 3-6. Analog Output Manual Calibration