3.5 ANALOG OUTPUT RECORDER OFFSET ADJUSTMENT
An offset can be added to individual output channels by pressing the following:
DIAG AIO RECORD OFFSET: 0 MV
+ 0 0 0 0 ENTR EXIT
DI AG
A
NALOG I / O CONFIGURATION
PREV NE XT ENTR EXIT
DIAG AIO
A
OUTS CALIBRATED: NO
SET> CAL EXIT
FROM ANALOG I/O CONFIGURATION MENU
DI AG AIO DATA_OUT_2 RANGE: 5V
SET> EDIT EXIT
These Keys set
the Recorder
Offset
of the s elected
channel
DIAG AIO DATA_OUT_2: 5V, NH3CNC1, CAL
< SET SET> EDIT EXIT
DIAG AIO DATA_OUT_2 REC OFS: 0 mV
< S ET SET> EDIT EXIT
Pressing ENTR records the new setting
and ret urns to t he previous menu
Pressing EXIT ignores
the new setting and
returns to the previous menu
Press SET> to select the Analog Out put
channel to be configured:
DISPLAYED AS = CHANNEL
DATA_OUT_1 = A1
DATA_OUT_2 = A2
DATA_OUT_3 = A3
DATA_OUT_4 = A4
Then Press EDIT to continue
Press SET Until
Figure 3-7. Analog Output Recorder Offset Calibration