Analyzer Operator Manual Configurable Analog Output Overview
05206H DCN5910 31
To calibrate the output channels as a group press the following, (Note: this is the
recommended calibration approach):
DIAG AIO AUTO CALIBRATING DATA_OUT_1
AUTO CALIBRATING DATA_OUT_2
AUTO CALIBRATING DATA
_
OUT
_
4
Exit at Any
Time to Return
to the main
DIAG Menu
DIAG
A
NALOG I / O CONFIGURATION
PREV NE XT ENTR EXIT
DIAG AIO
A
OUTS CALIBRATED: NO
< SET SET> CAL EXIT
If AutoCal has been manually
turned off for any channel the
message for that c hannel will be
similar to:
NOT AUTO CAL DATA_OUT_1
Exit to Return
to the I/O
Configuration
Menu
FROM ANALOG I/O CONFIGURATION MENU
DIAG AIO AOUTS CALIBRATED: YES
< SET SET> CAL EXIT
Figure 3-4. Analog Output Group Calibration