Teledyne 7300A Carbon Monoxide Alarm User Manual


 
4 Operation Model 7300A
4-18 Teledyne Analytical Instruments
4.4.1.2 Manual Mode Zeroing
Scroll to
Zero
and enter the
Zero
function. The screen that appears allows
you to select between automatic or manual zero calibration. Use the UP/DOWN
keys to toggle between AUTO and MAN zero settling. Stop when MANUAL
appears, blinking, on the display.
Select zero
mode: MANUAL
Enter
to begin the zero calibration. After a few seconds the first of three
zeroing screens appears. The number in the upper left hand corner is the first-
stage zero offset. The microprocessor samples the output at a predetermined
rate.
####.## % CO2
Zero adj:2048 CZero
The analyzer goes through CZero, FZero, and SZero. During CZero
and FZero, use the UP/DOWN keys to adjust displayed Zero adj: value as
close as possible to zero. Then, press
Enter
.
SZero starts. During SZero, the Microcontroller takes control as in Auto
Mode Zeroing, above. It calculates the differences between successive sam-
plings and displays the rate of change as Slope= a value in parts per million per
second (ppm/s).
####.## % CO2
Slope=#.### SZero
Generally, you have a good zero when Slope is less than 0.05 ppm/s for
about 30 seconds.
Once zero settling completes, the information is stored in the analyzers
memory, and the instrument automatically returns to the
Analyze
mode.
4.4.1.3 Cell Failure
Cell failure in the 7300A is usually associated with inability to zero the
instrument with a reasonable voltage differential across the Wheatstone bridge. If
this should ever happen, the 7300A system alarm trips, and the VFD displays a
failure message.
Cell cannot be balanced
Check your zero gas
Before replacing the sensor:
a. Check your zero gas to make sure it is within specifications.