Fluke 9150 Furnace User Manual


 
9150 Portable Furnace
User’s Guide
40
Command Description
Command
Format
Command
Example Returned
Returned
Example
Acceptable
Values
Display Temperature
Read current set-point s[etpoint] s set: 9999.99 {C or F} set: 150.0 C
Set current set-point to n s[etpoint]=n s=450 Instrument
Range
Set temperature units: u[nits]=c/f
CorF
Set temperature units to Celsius u[nits]=c u=c
Set temperature units to
Fahrenheit
u[nits]=f u=f
Read scan function sc[an] sc scan: {ON or OFF} scan: ON
Set scan function: sc[an]=on/of[f]
ON or OFF
Turn scan function on sc[an]=on sc=on
Turn scan function off sc[an]=of[f] sc=of
Read scan rate sr[ate] sr srat: 999.99 {C or F}/min srat: 10.0 C/min
Setscanrateto
n
degrees per
minute
sr[ate]=
n
sr=5 .1 to 99.9
Secondary Menu
Read proportional band setting pr[op-band] pr pb: 999.9 pb: 15.9
Set proportional band to
n
pr[op-band]=n pr=8.83 0.1 to 100
Read heater power
(duty cycle)
po[wer] po po: 999.9 po: 1
Ramp and Soak Menu
Read number of programmable
set-points
pn pn pn: 9 pn: 2
Set number of programmable
set-points to
n
pn=
n
pn=4 1 to 8
Read programmable set-point
number
n
ps
n
ps3 ps
n
: 9999.99 {C or F} ps1: 50.00 C
Set programmable set-point num
-
ber
n
to
n
ps
n
=
n
ps3=50 1 to 8, Instru
-
ment Range
Read program set-point soak
time
pt
n
pt3 ti
n
: 999 ti1: 5
Set program set-point soak time
to
n
minutes
pt
n
=
n
pt3=5 0 to 14400
Read program scan rate px
n
px3 sr
n
: 99.9 sr3: 11.3
Set program scan rate px
n
=
n
px3=10 .1 to 99.9
Read program control mode pc pc prog: {OFF or ON} prog: OFF
Set program control mode: pc=g[o]/s[top]/c[ont]
GO or STOP or
CONT
Start program pc=g[o] pc=g
Table 2 Communications Command Summary