200.0 C Well temperature in degrees Celsius
S
Access set-point memory
1. 200. Set-point memory 1, 200.0°C currently used
To change the set-point memory press “UP” or “DOWN”.
5. 900. New set-point memory 5, 900.0°C
Press “SET” to accept the new selection and access the set-point value.
S
Accept selected set-point memory
8.2.2 Set-point Value
The set-point value may be adjusted after selecting the set-point memory and
pressing “SET”.
0900.0 Set-point value in °C
The first digit flashes on and off. If the set-point value is correct, press “EXIT”
to display the temperature scale units.
If the first digit is correct press “SET”.
To adjust the first digit, press “UP” or “DOWN”. Then, press “SET” to go to
the second digit. The second digit of the temperature should now be flashing.
Adjust this digit by pressing “UP” or “DOWN”.
Press “SET” to accept the second digit and repeat until the last digit has been
adjusted.
0962.7 New set-point value
Press “SET” to accept the new set-point. If “EXIT” is pressed all changes made
to the set-point are discarded.
S
Accept new set-point value
8.2.3 Temperature Scale Units
Temperature Scale Units of the controller are set by the user to degrees Celsius
(°C) or Fahrenheit (°F). The units are used in displaying the well temperature,
set-point, and proportional band.
Press “SET” after adjusting the set-point value to change display units.
Un= C
23
8 Controller Operation
Temperature Set-point