Fluke 9150 Furnace User Manual


 
3 Specifications and Environmental
Conditions
3.1 Specifications
Temperature Range 150–1200°C (302–2192°F)
Display Resolution 0.1° to 999.9°, 1° above 1000°
Stability ±0.5°C
Display Accuracy ±5.0°C
Well Diameter 1.25" (32 mm)
Well Depth 4" (102 mm)
Heating Time 35 minutes to 1200°C
Cooling Time 140 minutes with block
Well to Well Uniformity ±0.5°C to ±1.0°C (insert “C” at 1200°C)
Stabilization 20 minutes
Power 115 VAC (±10%), 10.5 amps [230 VAC (±10%), 5.2 amps] switchable,
50/60 Hz, 1200 W
Size 12.4" H x 8.2" W x 12.4"D (315 x 208 x 315 mm)
Weight 28 lb. (13 kg)
Fault Protection Sensor burnout and short protection, over temperature thermal cutout
3.2 Environmental Conditions
Although the instrument has been designed for optimum durability and trou
-
ble-free operation, it must be handled with care. The instrument should not be
operated in an excessively dusty or dirty environment. Maintenance and clean
-
ing recommendations can be found in the Maintenance Section of this manual.
The instrument operates safely under the following conditions:
temperature range: 5 - 50°C (41 - 122°F)
ambient relative humidity: 15 - 50%
pressure: 75kPa - 106kPa
mains voltage within ± 10% of nominal
vibrations in the calibration environment should be minimized
altitude less than 2000 meters