Fluke 9150 Furnace User Manual


 
it can be adjusted.
00200 Program point 4 soak time set for 200 minutes
Press “SET” to save the new soak-time value or “EXIT” to discard changes
S
Accept the program point soak time
The next value to edit is the program scan rate. This value is ignored if scan is
not enabled for the unit (See Section 8.3.1).
Sr 4 Program point 4 scan rate
Press “SET” to edit the program scan rate.
S
Edit the program point scan rate
10.0 Current program point 4 scan rate
Use “UP” and “DOWN” to adjust the program scan rate.
11.3 New program point 4 scan rate
Press “SET” to save the new scan rate value or “EXIT” to discard changes.
S
Accept the program point scan rate
After “SET” is pressed the controller advances to the next program point or, if
there are no more program points to edit, exits to the Program Function Menu.
Repeat the above steps to edit any program point.
8.4.4 Program Function Mode
The next parameter is the program function or cycle mode. There are four pos
-
sible modes which determine whether the program scans up (from set-point 1
to n) only or both up and down (from set-point n to 1), and also whether the
program stops after one cycle or repeats the cycle indefinitely. The table below
shows the action of each of the four program mode settings.
Function Action
1 up-stop
2 up-down-stop
3 up-repeat
4 up-down-repeat
Pf=1
27
8 Controller Operation
Ramp and Soak Program