NGA 2000 CLD
Instruction Manual
HAS60E-IM-HW
June 2007
4 - 4 Maintenance and Service Emerson Process Management GmbH & Co.OHG
Figure 4-2: Converter Assembly
4-6 CONVERTER
HIGH TEMPERATURES !
While working at thermostated components inside
the Analyzer Modules hot components may be
accessible!
CAUTION
Refer to figure 4-1 and figure 4-2. To replace the con-
verter or temperature sensor, disconnect the two pneu-
matic tubes and two electrical connections. Unlace the
heater blanket, and remove the converter. Reassemble
in reverse order, ensuring that the converter is oriented
with the glass cloth at the bottom and the sensor is ori-
ented correctly inside the heater jacket.
Converter complete
(655250)
Heater Jacket
(655228-R1)
Temperature Sensor
(655282-R1)
Converter Tube
(655227)
Wrap with
Aluminium Foil
Connectors
(632784)