A SERVICE OF

logo

62 FONIX FP40 Portable Hearing Aid Analyzer
4.2.4 Viewing the results
See Figure 4.2.4 for an example of an ANSI 96 test sequence (including the telecoil measurement).
1. OSPL90: Output measurement at 90 dB SPL
2. RESP50 (or 60): Response measurement at 50 (or 60) db SPL at reference test gain
3. SPLITS: Telecoil response curve measurement
4. NR: Noise reduction used in tests
5. SPA/HFA: Frequencies used for the three frequency average
6. MAX: Maximum frequency response measured and the frequency at which it occurred
7. SPA/HFA OSPL90: Three frequency average of the OSPL90 curve
8. SPA/HFA FOG: Three frequency average of a full-on gain measurement at 50 (or 60) dB SPL
9. REFTG TARGET and MEASURED: The calculated reference test gain and the actual measured
reference test gain
10. EQ INP NOISE: Equivalent input noise
11. RESP LIMIT and F1 and F2: The response limit level and the two frequencies where the
response curve crossed over this level
12. THD: The total harmonic distortion measurements
13. HFA-SPLITS: The three frequency average of the telecoil SPLITS curve
14. STS-SPLITS: The difference between the high frequency average of the response curve and
the SPLITS curve
15. BATTERY: Battery current drain
16. I/O CURVES: Input/output measurements at up to five different frequencies
Figure 4.2.4—ANSI 96 results