Teledyne GFC 7001E Carbon Monoxide Alarm User Manual


 
Getting Started Model GFC7001E Carbon Dioxide Analyzer
Teledyne Analytical Instruments 55
3.6.2.4. Zero/Span Calibration
To perform the zero/span calibration procedure, press:
Press ENTR to changes
the OFFSET & SLOPE
values for the CO
measurements.
Press EXIT to leave the
calibration unchanged and
return to the previous
menu.
SAMPLE STABIL= XXXX PPM CO=XX.XX
< TST TST > CAL SETUP
M-P CAL STABIL= XXXX PPM CO=XX.XX
<TST TST> ZERO CONC EXIT
EXIT at this point
returns to the
SAMPLE menu.
Press ENTR to changes
the OFFSET & SLOPE
values for the CO
measurements.
Press EXIT to leave the
calibration unchanged and
return to the previous
menu.
SAMPLE RANGE=0.0 PPm CO= XX.XX
< TST TST > CAL SETUP
Set the Display to show
the STABIL test function.
This function calculates
the stability of the CO
measurement.
Toggle TST> button until ...
Allow zero gas to enter the sample port
at the rear of the analyzer.
Wait until STABIL
falls below 0.5 ppm.
This may take several
minutes.
SAMPLE STABIL= XXXX PPM CO=XX.XX
< TST TST > CAL SETUP
M-P CAL STABIL= XXXX PPM CO=XX.XX
<TST TST> ENTR CONC EXIT
M-P CAL STABIL= XXXX PPM CO=XX.XX
<TST TST> ZERO SPAN CONC EXIT
Allow span gas to enter the sample port
at the rear of the analyzer.
Wait until STABIL
falls below 0.5 PPM.
This may take several
minutes.
SAMPLE STABIL= XXXX PPM CO=XX.XX
< TST TST > CAL SETUP
M-P CAL STABIL= XXXX PPM CO=XX.XX
<TST TST> ENTR CONC EXIT
M-P CAL STABIL= XXXX PPM CO=XX.XX
<TST TST> ENTR CONC EXIT
The SPAN key now appears
during the transition from
zero to span.
You may see both keys.
If either the ZERO or SPAN
buttons fail to appear see the
Troubleshooting section for
tips.