Teledyne GFC 7001E Carbon Monoxide Alarm User Manual


 
Advanced Features Model GFC7001E Carbon Dioxide Analyzer
Teledyne Analytical Instruments 141
An alternative method for measuring the output of the Current Loop converter is to connect a 250 ohm 1%
resistor across the current loop output in lieu of the current meter (see Figure 3-7 for pin assignments and
diagram of the analog output connector). This allows the use of a voltmeter connected across the resistor to
measure converter output as VDC or mVDC.
V
+DC Gnd
Figure 7-7: Alternative Setup Using 250 Resistor for Checking Current Output Signal Levels
In this case, follow the procedure above but adjust the output for the following values:
Table 7-9: Current Loop Output Check
% FS
Voltage across
Resistor for 2-20 mA
Voltage across
Resistor for 4-20 mA
0 500 mVDC 1000 mVDC
100 5000 mVDC 5000 mVDC