Tektronix TDS524A Water System User Manual


 
Performance Tests
TDS 520A, 524A, 540A, & 544A Performance Verification
Test Record
Photocopy the next four pages and use them to record the performance test
results for your instrument.
TDS 500A Test Record
Instrument Serial Number: Certificate Number:
Temperature: RH %:
Date of Calibration: Technician:
Performance Test Minimum Incoming Outgoing Maximum
Offset Accuracy
CH1 Offset +1 mV
+100 mV
+1 V
– 1.6 mV
– 25 mV
250 mV
__________
__________
__________
__________
__________
__________
+ 1.6 mV
+ 25 mV
+ 250 mV
CH2 Offset +1 mV
+100 mV
+1 V
– 1.6 mV
– 25 mV
250 mV
__________
__________
__________
__________
__________
__________
+ 1.6 mV
+ 25 mV
+ 250 mV
CH3 Offset +1 mV
+100 mV
+1 V
– 1.6 mV
– 25 mV
250 mV
__________
__________
__________
__________
__________
__________
+ 1.6 mV
+ 25mV
+ 250 mV
CH4 Offset +1 mV
+100 mV
+1 V
– 1.6 mV
– 25 mV
250 mV
__________
__________
__________
__________
__________
__________
+ 1.6 mV
+ 25 mV
+ 250 mV
DC Voltage Measurement Accuracy (Averaged)
CH1 5 mV Vert scale setting,
–5 Div position setting
+ 1.0355 V __________ __________ + 1.0445 V
CH1 5 mV Vert scale setting,
+5 Div position setting
– 1.0445 V __________ __________ – 1.0355 V
CH1 200 mV Vert scale setting,
–5 Div position setting
+ 11.525 V __________ __________ + 11.675 V
CH1 200 mV Vert scale setting,
+5 Div position setting
– 11.675 V __________ __________ – 11.525 V
CH1 1 V Vert scale setting,
–5 Div position setting
+ 107.450 V __________ __________ + 108.550 V
CH1 1 V Vert scale setting,
+5 Div position setting
– 108.550 V __________ __________ – 107.450 V
CH2 5 mV Vert scale setting,
–5 Div position setting
+ 1.0355 V __________ __________ + 1.0445 V
CH2 5 mV Vert scale setting,
+5 Div position setting
– 1.0445 V __________ __________ – 1.0355 V