How to View or Change an HP-IB address from the Front Panel
Note
To set an HP-IB address from the front panel, the instrument
address on the rear panel HP-IB switch (Figure 3-2) must be set to
31 (all
1s).
1. Press SYSTEM
(e).
2.
Select HP-IB
Menu
Adrs
Nenu.
3.
The synthesizer displays the three address softkeys: 8360
A&s,
Meter Adrs , and Printer Adrs .
4. Select the desired softkey.
5. The synthesizer displays the address selected for that instrument.
6. If you want to change the address, use the keypad to enter the
desired address (0 to 30), then press
@%?j.
Remember
If the synthesizer displays Rear panel HP-IB address must be 31
(11111) in order to change current address
(=Xx>,
the address
on the rear panel HP-IB switch (Figure 3-2) is set to something other
than 31 (all
1s).
How to Prevent a Front Panel Change to an HP-IB Address
To disable the address softkeys, set the instrument address on the
rear panel HP-IB switch (Figure 3-2) to any address other than 31
(all
1s).
How to Set the HP-IB Address on a Synthesizer without a Front Panel
If your synthesizer does not have a front panel, set the address on
the rear panel HP-IB switch (Figure 3-2) to the address you want
(factory default is 19).
Mating Connectors
All of the externally mounted connectors on the instrument are
discussed in the “Operating and Programming Reference” chapter,
under “Connectors.”
If you are interested in the HP part number for
a connector, see “Replaceable Parts” in this manual set.
10 MHz Frequency
To keep the internal
timebase
frequency reference oven at operating
Reference Selection
temperature, the synthesizer must be connected to ac line power.
and Warmup Time
The synthesizer requires approximately 30 minutes to warm up
from a cold start before the OVEN display message goes off. With
a stable outside temperature, internal temperature equilibrium is
reached after approximately two hours. For additional information on
warmup times, see “Specifications,” in this volume.
3-8 INSTALLATION
Specifications