HP (Hewlett-Packard) 24A Vacuum Cleaner User Manual


 
Leveling Point
Module
Function Group
Menu Map
Description
Programming Codes
See Also
ALC
1
This
softkey
lets you set the synthesizer to level at the output of an
HP
8355X
series millimeter-wave source module. All models of the
HP 8360 series synthesized sweepers drive mm-wave source modules.
High power models of HP 8360 drive the mm-wave source modules
directly and to specified power levels. An HP
8349B
power amplifier
is needed in other configurations. The source module interface
multi-pin connector provides the communication path between the
synthesizer and mm-wave source module. An asterisk next to the key
label indicates that this feature is active.
SCPI: POWer:ALC[:SOURce]
MMHead
Analyzer: SHA2
(ALC),
CONNECTORS
“Externally Leveling the Synthesizer,” in Chapter 1.
Leveling Point
PwrMtr
Function Group
Menu Map
Description
Programming Codes
See Also
ALC
1
This softkey lets you set the synthesizer to level at the power
sensor of an external power meter. This mode of operation requires
a feedback connection from the power meter to the EXT ALC
BNC located on the synthesizer. An asterisk next to the key label
indicates that this feature is active.
SCPI: POWer:ALC[:SOURce]
PMETer
Analyzer: A3
(ALC),
CONNECTORS
“Externally Leveling the Synthesizer,” in Chapter 1.
L-4
Operating and Programming Reference
HP 8360
User’s Handbook