O2 Innovations pmn Carbon Monoxide Alarm User Manual


 
TARGET Series ENMET Corporation
4
2.4 Locations and Data logging
The TARGET instrument has the ability to store users or location information and log exposure information. TARGET
Trakker software is required for these operations. Further details on its operation and interaction with the instrument is
provided in software manual. If desired, contact ENMET or your local distributor.
3.0 Instrument Operation
3.1 Turning the Instrument On and Off
3.1.1Turning the TARGET On
Press and hold the ON/OFF button. The audio alarm may squeal slightly while holding the ON/OFF button. This is
normal. When the instrument turns on, it beeps once and all of the alarm LEDs light up. The TARGET performs a one
minute warm-up procedure which includes a self test, checking sensor signals and instrument electronics. At the end
of the one minute warm-up procedure, the instrument will beep several times before giving the user the option to
perform an auto zero function, or other functions that appear in the Pop-Up menu. See section 3.5 for further
information.
Example: Warm-up Displays
3.1.2 Turning the TARGET Off
Press and hold the ON/OFF button for approximately three seconds. The instrument shows a three second count
down and beeps just before shutting down. To protect the user from accidental shut down, the ON/OFF button must
be held for the duration of the three second count down. The audio alarm beeps once to indicate instrument shut
down.
Example: Power Down Display
ENMET CORP
TARGET Series
GAS MONITOR
Rev:3.0-3.1
02 20.9%
CO 0ppm H2S 0ppm
Mth0%LEL BRH 0ppm
Warming up...30
POWER DOWN SEQUENCE
Press and hold power
switch for 3sec
2