button at the Exit line to go back to the Maintenance menu. STEL and TWA alarms are preset at the factory and can
only be changed using the TARGET Trakker software package.
SELECT
Example: Display of Set Alarm Menu
6.4.3 Enable Sensors
From the Advanced Maintenance menu, press the MENU button twice once to put the asterisk next to the Enable
Sensors line. Press the SELECT button to enter the Enable Sensors screen. Use the MENU button to move the
asterisk next to the gas whose alarm needs to be changed. Press the SELECT button to toggle the sensor ON or OFF.
Press SELECT at the Exit line to go back to the Maintenance menu.
SELECT
6.4.4 Set Clock/Chirp
From the Advanced Maintenance menu, press the MENU button three times to put the asterisk next to the Set
Clock/Chirp line. Press the SELECT button to enter the Set Clock/Chirp screen. Use the MENU button to move
the asterisk next to the value that needs to be changed. Press the SELECT button to move a cursor to the first digit.
The MENU button is used to increment the digit and the SELECT button to move to the next digit.
Time must be entered in a 24 hour format (i.e. 4:30 pm is entered as 16:30). The confidence chirp sounds at interval
entered in this screen. Factory default value is 120 seconds.
Press the SELECT button at the Exit line to go back to the Maintenance menu.
SELECT
Example: Display while Setting Clock and Chirp interval
6.4.5 Set Password
From the Advanced Maintenance menu, press the MENU button four times to put the asterisk next to the Set
Password line. Press the SELECT button to enter the Set Password screen. Use the MENU button to increment the
digit and the SELECT button to move to the next digit. Any four digit password from 0000 to 9999 may be entered
by this operation.
SELECT
Example: Display Changing Password
SET ALARMS
*O2 19.5 / 23.5 %
CO 35 ppm
H2S 10 ppm
Calibrate
*Set Alarms
Enable Sensors
Set Clock/Chirp
Calibrate
Set Alarms
*Enable Sensors
Set Clock/Chirp
ENABLE SENSORS
*O2 ON
CO ON
H2S ON
Calibrate
Set Alarms
Enable Sensors
*Set Clock/Chirp
*M/D/Y: 05/22/99
Time (24Hr): 18:53
Chirp (sec): 120
Exit
*Set Password
Set Battery
Set Op. Menu Type
Exit
CHANGE PASSWORD
Enter new value:
1270