Conductivity Probe
8990options.fm
Section 5
c. If the calibration temperature is lower than the labeled value, then
subtract that value from the standard (1.0 mS) to get the actual value
to be used for calibration.
1.0 mS - (correction factor) 0.13002 = 0.86998 mS
d. If the calibration temperature is higher than the labeled value, then
add that value to the standard (1.0 mS) to get the actual value to be
used for calibration.
7. Using the value that was calculated in step 6, enter the conductivity of the
solution then press
ACCEPT to complete the calibration.
Calibrating the Conductivity Temperature
Note: Conductivity temperature
calibration is only necessary when
logging temperature.
1. Place the probe in a liquid and wait for the temperature reading
to stabilize.
2. Enter the actual temperature of the liquid (the current reading is shown for
reference). Temperature calibration is complete.
Table 12 Conductivity Values at Temperature for Hach KCl Solution
Solution
Temp °C
Calibration Value
to be Entered
Solution
Temp °C
Calibration Value
to be Entered
Solution
Temp °C
Calibration Value
to be Entered
30 1.099 25 1.000 20 0.902
29.8 1.095 24.8 0.996 19.8 0.898
29.6 1.091 24.6 0.992 19.6 0.894
29.4 1.087 24.4 0.988 19.4 0.890
29.2 1.083 24.2 0.984 19.2 0.886
29 1.079 24 0.980 19 0.882
28.8 1.075 23.8 0.976 18.8 0.878
28.6 1.071 23.6 0.972 18.6 0.874
28.4 1.067 23.4 0.968 18.4 0.870
28.2 1.063 23.2 0.965 18.2 0.866
28 1.059 23 0.961 18 0.862
27.8 1.055 22.8 0.957 17.8 0.858
27.6 1.051 22.6 0.953 17.6 0.854
27.4 1.047 22.4 0.949 17.4 0.850
27.2 1.043 22.2 0.945 17.2 0.846
27 1.039 22 0.941 17 0.842
26.8 1.035 21.8 0.937 16.8 0.838
26.6 1.032 21.6 0.933 16.6 0.835
26.4 1.028 21.4 0.929 16.4 0.831
26.2 1.024 21.2 0.925 16.2 0.827
26 1.020 21 0.921 16 0.823
25.8 1.016 20.8 0.917 15.8 0.819
25.6 1.012 20.6 0.913 15.6 0.815
25.4 1.008 20.4 0.909 15.4 0.811
25.2 1.004 20.2 0.905 15.2 0.807