Hach 900 MAX Water Heater User Manual


 
Section 3
Page 53
8990startup.fm Advanced Sampling
Bottle Number
If the Program Complete Output is disabled, then it is used in conjunction with
this Special Output to transmit the bottle number to the connected device.
The Special Output signal can be configured to activate during one of the
following conditions:
After Each Sample—One second pulse at the completion of each sample
cycle. It signals an external datalogger or PC that a sample cycle was
initiated. When this choice is enabled, the sample success/failure is also
transmitted to the external datalogger by means of Pin F on the
Auxiliary connector.
Only When Pumping—During sample intake portion of cycle only,
ignoring all purges and rinse cycles. Used for tipping solenoid or ball
valves when sampling from a pressurized line.
From Rinse To Purge—During entire sample cycle, including all purge
and rinse cycles. Used for tripping solenoid or ball valves when sampling
from a pressurized line.
Step 15 - Start/Stop Times
Start/Stop Times start and stop a program at pre-arranged times. For
example, this feature can be used to run a program during weekdays and stop
the program during weekends, restarting on the following Monday. It can also
halt sampling for nighttime shutdowns.
Up to 12 Start and 12 Stop time entries can be set in any one program. Either
a Start Time and Date or a Start Time and Day of the Week may be selected.
15-A. Highlight the Start/Stop Time choices on the Advanced Sampling menu.
Press
SELECT to continue.
15-B. Enable or Disable Start/Stop Times using the
CHANGE CHOICE key.
Press
ACCEPT to continue.
15-C. Press
CHANGE CHOICE to select either Time/Date or Time/Weekday.
Select Time/Date if the start and stop times are longer than one week.
Select Time/Weekday if the program repeats on a daily or weekly basis.
15-D. Press
ACCEPT to continue.
11:00 AM 21 - APR - 01 ADVANCED SAMPLING
SELECT
PGM CMPLT OUTPUT
SETPOINT SAMPLING
SPECIAL OUTPUT
START/STOP TIMES
RETURN
STORM WATER
TIMED BOTTLE SETS