Advanced Sampling
8990startup.fm
Section 3
Step 19 - Variable Intervals
19-A. Highlight Variable Intervals using the UP and DOWN keys on the
Advanced Sampling Menu. Press
SELECT to continue.
19-B. Enable or Disable Variable Intervals using the
CHANGE CHOICE key.
Press
ACCEPT to continue.
19-C. Set sample intervals using the numeric keypad. Press
ACCEPT to enter
another interval or press
ACCEPT AS FINAL to return to the Advanced
Sampling menu and continue to Variable Volumes.
Step 20 - Variable Volume
20-A. Highlight Variable Volume using the UP and DOWN keys on the
Advanced Sampling Menu. Press
SELECT to continue.
20-B. Enable or Disable Variable Volume using the
CHANGE CHOICE key.
Press
ACCEPT to continue.
20-C. Enter the Sample Volume counts using the numeric keypad. Press
ACCEPT to return to the Advanced Sampling Menu.
11:00 AM 21 - APR - 01 ADVANCED SAMPLING
SELECT
STORM WATER
TIMED BOTTLE SETS
UPSET SAMPLING
VARIABLE INTERVALS
RETURN
VARIABLE VOLUME
11:00 AM 21 - APR - 01 ADVANCED SAMPLING
SELECT
TIMED BOTTLE SETS
UPSET SAMPLES
VARIABLE INTERVALS
VARIABLE VOLUME
RETURN