Eagle Home Products Eagle Series Smoke Alarm User Manual


 
Eagle Instruction Manual Calibration 41
Calibration
Calibrate the Eagle when you replace a sensor. Also calibrate the Eagle
periodically to assure proper sensor response.
You can program the Eagle to notify you when it is due for calibration
(see “Updating the Time Calibration Setting” on page 38). The
frequency of calibration depends upon the amount and type of use. A
typical calibration frequency is once per month.
Calibration Supplies and Equipment
To calibrate the Eagle, you need:
Known calibrating samples of combustible and the appropriate
toxic gases. The samples should have concentrations in
approximately the middle of the range of detection.
An oxygen-free source, such as 100% nitrogen or CO in a nitrogen
balance
A demand-flow regulator to provide adequate sample gas flow
For one-source, auto-calibration, you can use the RKI Four-Gas
Cylinder to adjust all the sensors at the same time, automatically, with
no need for a zero-oxygen source. This section includes instructions
for calibration with the demand-flow regulator and RKI Four-Gas
Cylinder. This section also includes instructions for calibration with
individual gas sources.
Preparing for Calibration
1. Take the Eagle to a non-hazardous location with fresh-air
conditions.
2. Turn on the Eagle and allow one minute for warm up.
3. Press and hold the AIR/ button until a tone sounds.
The Eagle automatically sets the combustible gas and toxics
circuits to zero and the oxygen circuit to 20.9%.
4. Screw the regulator to the calibration cylinder.
5. Connect the calibration tubing to the regulator.