Eagle Instruction Manual Setup Mode 35
Turning the User ID Function On or Off
With User ID Input ON, the User and Station ID screen displays
during start up. From this screen, you can enter user, location, or other
information at the beginning of each gas detection session (see
page 19).
With User ID Input OFF (default), the User and Station ID screen
does not display during start up.
1. From the main menu, select the USER ID menu option.
2. Press the AIR/ or SHIFT/ button to display the desired
setting.
3. Press the POWER/ENTER button to enter the setting and return
to the main menu.
Updating the Auto Calibration Settings
The Eagle stores default calibration settings. This allows you to
calibrate all Eagle channels simultaneously with a calibration cylinder
that contains all required target gases (for example the RKI four-gas
calibration cylinder).
The Eagle includes default auto calibration settings for most target
gases. For gases without default auto calibration, the setting is 0.
NOTE: You can also update auto calibration settings in Calibration
mode. If you update auto calibration settings in Calibration
mode, you must continue with the calibration procedure.
Updating these settings in Setup mode allows you to update
the settings without calibrating the sensors.
1. From the main menu, select the AUTO CALIBRATION menu
option. (To display the combustible gas channel in PPM, press the
LEL/PPM button.)