Teledyne 6750 Carbon Monoxide Alarm User Manual


 
Total Organic Carbon Analyzer Maintenance
Teledyne Analytical Instruments 65
6. Remove 4 pump mounting screws
for pump removal.
7. Replace motor and reassemble in
reverse step sequence.
Figure 5-14: Removing the Pump
Note: Actual mounting panel may appear somewhat different but
mounting hardware is the same.
5.3.12.2 PUMP HEAD TUBING REPLACEMENT
1. Separate the end bells (the pump head
halves).
2. Hold the end bell containing the rotor as
shown with the tubing retainer grooves
pointing down.