Teledyne 6750 Carbon Monoxide Alarm User Manual


 
Maintenance Model 6750
Teledyne Analytical Instruments 44
From this screen the operator may perform individual
testing of components related to functionality and
settings.
5.2 Troubleshooting
If you are experiencing trouble with your instrument, refer to Table
5-1: Troubleshooting.
Table 5-1: Troubleshooting
PROBLEMS
SOLUTION
Analyzer will not turn on
Check fuse & power connections.
Look for loose or damaged wiring.
Check for 12V DC from power supply.
Check power supply fuse.
Analyzer does not Calibrate, is
sluggish
Look for & correct any liquid or gas line
leaks.
Analyzer is excessively noisy
Check for leaks at tubing connections
Display not functioning
Check cable connections
Check power supply voltage
Suspect:
Malfunctioning Computer (P/N ST13039).
Master Control Panel Assembly (P/N
ST13042-1)
Pump head not rotating
Refer to section “Pumps” 12.3.12
Verify that carrier loss did not cause the
pumps to be turned off.
Check for AC voltage at terminal strip on
inside of application panel.
Pump head rotating but not
pumping fluid
Refer to section “PUMPS”
Change tubing
Verify sample line is in liquid.