Teledyne 6750 Carbon Monoxide Alarm User Manual


 
Setup and Operation Model 6750
Teledyne Analytical Instruments 34
From this menu verify or change the concentration alarm setpoints
using the onscreen scroll arrows.
4.2.4 Stream Sequencer
If your analyzer has been configured for a multi-stream sequencer
(P/N ST200009), from the [Edit] Menu, select [Sequencer]. The
following menu appears:
From this menu, you can:
Configure the duration of analysis for each stream (how long
it will analyze that stream before selecting the next stream)
Configure the changeover time (to allow the prior stream to
pass through sampling systems and clear the analyzer before
analysis of the next steam begins),
Enable and disable each stream analysis (for example, to
concentrate on only one stream).
The following window is illustrative of a dual stream [Run] screen.