Total Organic Carbon Analyzer Setup and Operation
Teledyne Analytical Instruments 31
Figure 4-2: Gas Liquid Separator D. I. Filling
After reagents and DI have filled all tubing lines, it is
recommended to proceed with IR Calibration followed by cal (liquid
calibration) of the system. Then the analyzer may be placed in service.
4.2 Menus
The advanced design of the Model 6750 TOC Analyzer eliminates
complicated, routine, and sometimes confusing menus. The
operator/software interface is simple and easy to use. Menus prompt the
user for required actions or input.
The analyzer has no manual adjustments. All calibrations and
operations are computer controlled by the operator following menu
prompting and selection of the operation of choice.
The following menu and operation descriptions are intended to
guide the operator through all the functions of the analyzer.
After primary power (110/220VAC) has been applied to the
analyzer and self-diagnostic procedure has been completed, the system
automatically boots up to the following RUN Screen. If not, turn main
power OFF, then ON to reboot.
4.2.1 Run Mode
RUN – This is the normal analysis mode. In this mode, the display
indicates: