Sanyo UV Carbon Monoxide Alarm User Manual


 
136
45
12. Dilute the remaining H
2
O
2
reagent in the H
2
O
2
Generator with a large volume of water and dispose of
it. Rinse and wash the H
2
O
2
Generator with distilled water. Then keep the H
2
O
2
Generator in a clean
environment outside of the chamber.
Note: After H
2
O
2
decontamination, cover the connector on the chamber side with the connector cap.
13. After H
2
O
2
decontamination, surplus H
2
O
2
liquid will remain at the bottom of the chamber and in the
bottom part of the H
2
O
2
Generator duct. This solution contains H
2
O
2
at a low concentration (about 1% or
less), so put on protective glasses and rubber gloves and wipe it up with a non-woven cloth.
14. Ventilate the chamber sufficiently and place all the attachments back into the chamber.
Note: Be sure to attach the connector cap.
CAUTION
The electric lock will remained locked if power fails during H
2
O
2
decontamination. After the power comes
back ON, the H
2
O
2
gas resolution process will start execution and finish automatically. Execute the
decontamination again because the decontamination will be incomplete.
WARNING
Never open the door by unlocking it with the unlock key during H
2
O
2
decontamination or H
2
O
2
gas
resolution with the UV lamp. H
2
O
2
gas leakage is potentially harmful to health.
Precautions when handling H
2
O
2
decontamination reagent
Observe the following precautions when handling the H
2
O
2
Decontamination Reagent (MCO-H2O2).
z
Handling Precautions
x Wear protective equipment, such as protective glasses and rubber gloves.
x Do not use fire in the area where the reagent is being handled.
x Do not leave any reagent in the container after it has been used or while it is being used.
x Do not place inflammable or combustible materials near the area where the reagent is handled.
z
Precautions for Storage
x Always close the container cover securely to prevent impurities from becoming mixed in the reagent.
x Check the container to make sure that there is no damage, corrosion, or cracking.
x Store the container with the inlet facing upwards, and make sure that the container will not tip or be
knocked over.
x If possible, store the reagent in a location where water is available and where the floor can be easily
washed.
z
Precautions for Disposal
x Dispose according to the rule in your country.