Sanyo UV Carbon Monoxide Alarm User Manual


 
129
38
12. Clean all the attachments with a diluted neutral detergent, and then rinse them thoroughly with
distilled water.
13. Wipe the trays, the inner attachments such as the chamber circulation fan, and the chamber sides
with sterilizing alcohol. Be careful not to leave any residual alcohol.
14. Wipe the water level sensor with sterilizing alcohol. Be careful not to leave any residual alcohol. When
cleaning the sensor, take care not to apply excessive force to the lead wires.
15. To reinstall all the attachments, perform the procedure in reverse order from step 14.
Note:
• When installing the chamber circulation fan, insert
it securely on the motor shaft. Lightly turn the fan
manually to make sure that it does not strike the rear
panel. Improper insertion may cause poor
performance.
• As shown in Fig. 11, set the tray with only the front
edge bent down. If the tray is set in the wrong
direction, it may not be level and may become
unstable.
Bent down
Fig. 11