111
20
BASIC PARAMETERS
㩷
㩷 Setting the chamber temperature and CO
2
density㩷
㩷
The setting procedure for the chamber temperature and CO
2
density are given below. (Default settings:㩷
Chamber temperature: 37
o
C, CO
2
density: 0%)㩷
The Incubator automatically starts operation using these settings after the power is turned ON.
㩷
1. From the Top Display, press the MENU Key to access the menu. Select Set and press the ENTER
Key.
㩷
㩷
㩷
㩷
㩷
㩷
㩷
㩷
㩷
㩷
㩷
㩷
2. The Stand-By Setting Display will be displayed. Set the parameters.㩷
㩷
㩷
㩷
㩷
㩷
㩷
㩷
㩷
㩷
㩷
㩷
㩷
The setting ranges of the parameters are as follows:
ƔTemperature: 0 to 50
o
C. To 37.0
o
C, enter 370.
ƔCO
2
density: 0 to 20%. To set 5.0%, enter 050.㩷
㩷
3. Press the MENU Key to display the menu after setting the parameters. Select OK and press the
ENTER Key. The settings will be saved.
㩷
Note:
When starting the Incubator for the first time or after not using it for an extended period of time, allow at
least 4 hours for the chamber temperature, humidity, and CO
2
sensor to become stable after setting the
desired chamber temperature at a 0% CO
2
density. Then change the setting to the desired CO
2
density.
㩷
㩷
㩷
1
S t a n d - by S e t t ing
2
T e m p e r a t u r e
3 7 . 0
o
C ( 0. 0
o
C- 50. 0
o
C )
3
C O 2 D e n s i t y 5 . 0 %
(0.0% - 20. 0 % )
4
5
K e y L o c k 0
( 0 . U n l o c k 1 . L o c k )
6
H i g h L i m i t 5 2 . 0
o
C
M E N U
O K
C a n c e l
1
T e
m p 3 7 . 0
o
C C O 2 5 . 0 %
2
3
4
5
S t a t u s
6
O K
7
㧟㧣㧜
㧡㧜
M E N U
S e t
L o g
T o o l s