Emerson Process Management 2500 Home Security System User Manual


 
Mounting multiple transmittersFigure 3-2:
A
B
A. 0.33 in or greater (8.5 mm or greater)
B. End bracket or end stop; 0.33 in (8.5 mm) minimum spacing
3.2
Mount the remote core processor
This procedure is required only for remote core processor with remote transmitter
installations.
For mounting the remote core processor to a wall:
Use four 5/16-inch diameter (or M8) bolts and nuts that can withstand the process
environment. Micro Motion does not supply bolts or nuts (appropriate bolts and
nuts are available as an option).
Ensure that the surface is flat and rigid, does not vibrate, or move excessively.
For mounting the remote core processor to an instrument pole:
Use two 5/16-inch U-bolts for 2-inch pipe, and four matching nuts, that can
withstand the process environment. Micro Motion does not supply U-bolts or nuts.
Ensure the instrument pole extends at least 12 inches (305 mm) from a rigid base,
and is no more than 2 inches (50.8 mm) in diameter.
1. If desired, reorient the core processor housing on the bracket.
a. Loosen each of the four cap screws (4 mm).
b. Rotate the bracket so that the core processor is oriented as desired.
c. Tighten the cap screws, torquing to 30 to 38 in-lbs (3 to 4 N-m).
Mounting and sensor wiring for remote core processor with remote sensor installations
Installation Manual 13