2 Instruction 3015-4286
Safety Precautions
AC Power Supply
The AGMSZ uses a universal power supply that is capable of accepting inputs of 100 to 240 VAC,
50/60 Hz. The monitor’s power consumption is 15 Watts. It is highly suggested that the monitor be
connected directly to the AC power source, preferably on its own circuit (with UPS or surge protection).
A switch or circuit breaker rated 1.0 A, 250 VAC, with a minimum terminal spacing of 3.0 mm must be
attached to the monitor’s AC power leads. This switch must also be located in close proximity to the
monitor, and be in easy reach of the operator. This switch should also be clearly marked as the monitor’s
main AC disconnect device.
Protective Grounding
Under no circumstances should the AGMSZ be operated without connection to a protective ground. Doing
so poses a potential shock hazard and is also a violation of electrical safety standards applicable to this
type of equipment.
Explosive Atmosphere
Do not operate this equipment in the presence of flammable liquids, vapors or aerosols. Operation of any
electrical equipment in such an environment constitutes a safety hazard.
Proper Exhaust Venting
It is imperative that the monitor’s exhaust port be vented as described in this manual. Failure to do so
may constitute a safety hazard.
Working Inside Monitor
Extreme care should be exercised when accessing the interior of the monitor. Only qualified electrical
maintenance personnel should perform connections and adjustments. Always remove AC power before
working inside the monitor.
Misuse and Modifications to Monitor
The protection provided by the monitor may be impaired if the monitor is used in a manner not specified by
Bacharach, Inc. Changes or modifications to this monitor, not expressly approved, will void the warranty.
In Case of Malfunction
Do not continue to use this equipment if there are any symptoms of malfunction or failure. In the case of
such occurrence, de-energize the power supply and contact a qualified repair technician or the nearest
Bacharach Service Center. Use ONLY the provided knockouts for electrical and communication wiring.
Drilling into the box will void the warranty.
Fusing
F1, F2: 1.0 A, 250 V, Type “F”
Installation Category
Installation Category II, Pollution Degree II, as defined by UL.
Altitude Limit
6,562 ft (2,000 m)
Cleaning
To clean the outside of the case, DO NOT use soap and water. USE a dry cloth.