Bacharach 3015-4286 Carbon Monoxide Alarm User Manual


 
6 Instruction 3015-4286
Mounting Instructions
The AGMSZ should be installed plumb and level and securely fastened to a rigid mounting surface.
The enclosure utilizes keyhole mounting brackets designed for #12 pan head screws. Locate the four
screws as shown in the diagram below and leave the screw heads protruding approximately 3/16".
Enclosure Mounting Specs
Hold the monitor flat against the mounting surface and allow it to slide down engaging the screw heads in
the keyhole slots of the mounting brackets. Adjust the screws as necessary to hold the monitor securely
against the mounting surface.
Connecting Gas Sample, Purge, and Exhaust Lines
Overview
A single gas-sample line needs to be run from the AGMSZ to the area of the facility to be monitored. An
optional purge line can be run to bring fresh (zero) air into the monitor for detector zeroing purposes. And
an optional exhaust line can be installed to vent the sample gas away from the monitor. All tubing
connections are made on the bottom of the enclosure as shown in the illustration below.
AGMSZ Bottom View
GAS SAMPLE
EXHAUST PORT
GAS SAMPLE
INLET PORT
½" / ¾" SERVICE
KNOCKOUT*
PURGE AIR
INLET PORT
½" / ¾" SERVICE
KNOCKOUT*
* There is an additional Service Knockout on top of the case