Agilent Technologies 6814B Air Conditioner User Manual


 
Front Panel Operation - 4
35
Scrolling Keys
Scrolling keys let you move through the commands in the presently selected function menu.
These scroll keys let you move through the choices in a command list.
Press to bring up the next command in the list. Press to go back
to the previous command in the list. Function menus are circular; you can
return to the starting position by continuously pressing either key. The
following example shows the commands in the Input function menu:
These shifted scroll keys apply only to the Harmonic and List functions.
Press these keys to step through integers 0 through 50 when specifying the
desired harmonic number, or 0 through 99 when specifying the desired list
point. Hold down these keys to rapidly access any harmonic or list point.
These Entry keys let you scroll through choices in a parameter list that apply
to a specific command. Parameter lists are circular; you can return to the
starting position by continuously pressing either key. If the command has a
numeric range, these keys increment or decrement the existing value.
Meter Display Keys
Metering keys control the metering functions of the ac source.
Press this key to access the meter menu list.
Display Measurement
<reading>V <reading>Hz rms voltage and frequency (the default)
<reading>V <reading>A rms voltage and rms current
<reading>A <reading>Hz rms current and frequency (the default)
<reading>V <reading>W rms voltage and power
<reading> CREST F current crest factor
<reading>A PK REP peak current, repetitive
<reading>A PK NR peak current, nonrepetitive
1
<reading> VA apparent power
<reading> VAR reactive power
<reading>W TOTAL total power of all phases
2
<reading> PFACTOR power factor
<reading>A NEUTRAL neutral rms current
2
Press this key to specify the following metering functions.
Display Command Function
INP:COUP <char> Choose meter coupling (AC, DC or ACDC)
WINDOW <char> Select harmonic measurement window meter
(KBESSEL, RECT)
Notes:
1
Displays the highest peak current since it was last cleared. The value is cleared when
you scroll into this selection or press or
2
This selection only valid for Agilent 6834B
reading = the returned measurement
char = a character string parameter
and scroll through the command list.
and scroll through the parameter list.
Index
Shift
Shift
Index
¯ °
Meter
Input
Enter
Clear Entry