Teledyne 300M Carbon Monoxide Alarm User Manual


 
Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A
Table 4-1: Setup Functions (Continued)
Button Sequence Function Default Limits
SETUP-MORE-DIAG-
SIGNAL I/O
View the state of
internal signals
N/A N/A
SETUP-MORE-DIAG-
ANALOG OUTPUT
Generate Analog
Output Test Pattern
N/A N/A
SETUP-MORE-DIAG-
D/A CALIBRATION
Calibrate D/A and A/D
Converters
N/A N/A
SETUP-MORE-DIAG-
ELECTRICAL TEST
Generate Electrical
Test Output
N/A N/A
SETUP-MORE-DIAG-
DARK
CALIBRATION
Adjust Dark Offset 125mV 75 - 175 mV
SETUP-MORE-DIAG-
FLOW
CALIBRATION
Calibrate sample
flowrate
800 cc/m 800 cc/m
SETUP-MORE-
DIAG- TEST CHAN
OUTPUT
Select TEST to Analog
Output
None
SETUP-MORE-DIAG-
RS-232 OUTPUT
Generate RS-232
Output Test Pattern
N/A N/A
SETUP-MORE-
ALRM-ALM1
Enable alarm 1 and set
limit
ON 100 PPM
SETUP-MORE-
ALRM-ALM2
Enable alarm 2 and set
limit
ON 300 PPM