Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A
Table 4-1: Setup Functions
Button Sequence Function Default Limits
SETUP-CFG-NEXT List Software
Configuration
CO Mach N/A
SETUP-ACAL-MODE Define/Change
AutoCal Sequences
Disabled Zero, Span, Zero-
Span
SETUP-DAS-EDIT Define/Change DAS
Data Channels
SETUP-DAS-VIEW View DAS Data N/A N/A
SETUP-RNGE-MODE Set Range Mode Single Single, Dual, Auto
SETUP-RNGE-SET Set D/A output range 500 PPM 1-20,000 PPM
SETUP-RNGE-UNIT Set Measurement Units PPM PPM. mg/m
3
, ug/m
3
SETUP-PASS Password Enable ON OFF-ON
SETUP-CLK-TIME Set Time-of-Day 00:00 00:00-23:59
SETUP-CLK-DATE Set Current Date 01 JAN 00 31 DEC 99
SETUP-MORE-
COMM-BAUD
RS-232 baud rate 2400 baud 300, 1200, 2400,
4800, 9600, 19.2
SETUP-MORE-
COMM-ID
Analyzer ID number 0000 0000-9999
SETUP-MORE-
VARS-
DAS_HOLD_OFF
Set Hold-Off Interval 15 min 1 - 60 min
SETUP-MORE-
VARS- DYN_ZERO
Enable Remote
Dynamic Zero
Adjustment
OFF OFF-ON
SETUP-MORE-
VARS- DYN_SPAN
Enable Remote
Dynamic Span
Adjustment
OFF OFF-ON
SETUP-MORE-
VARS- RS232_MODE
Set RS-232 Mode 8
SETUP-MORE-
VARS- CLOCK_ADJ
Set Clock Adjustment
Rate
0 -15 to +15
4-8