Teledyne 300M Carbon Monoxide Alarm User Manual


 
Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A
Table 1-2: Final Test and Calibration Values
Test Values
Observed
Value
Units Nominal Range
CO MEAS mV 2500 - 4700
CO REF mV 2500 - 4700
MR RATIO 1.1 – 1.22
SAMPLE PRESS in-Hg 27 - 30
VACUUM PRESS in-Hg
10 ± 5
SAMPLE FLOW CC/MIN 700 - 900
BENCH TEMP
°C 48 ± 1
DC POWER
SUPPLY
mV
2500 ± 200
SAMPLE TEMP
°C
45-50
BOX TEMP
°C
8- 40
WHEEL TEMP
°C 68 ± 2
Diagnostic
Dark Ref mV
125 ± 50
Dark Meas mV
125 ± 50
Electrical test PPM
400 ± 4
Span and CAL Values
CO Span Conc PPM 400
CO Slope
1.0 ± 0.3
CO Offset -0.1 to +0.1
Noise at Zero (p-p) PPM 0.5
Noise at Span (p-p) PPM <1.0% of reading
Factory Installed Options Option Installed
Factory Installed Options Option Installed
Power Voltage/Frequency
Rack Mount, w/ Slides
Internal Zero Span Valves
Analog Voltage Range 0 - V
PROM # Serial #
Date Technician
1-16