SPX Cooling Technologies TG MAG15-50 Water Pump User Manual


 
43
A.0500.551 – IM-TGMAG/02.00 EN (02/2008)
Symptom Cause Remedy
Pump leak O-rings, sealing rings or gasket damaged
or not correctly assembled
41 Check condition of O-rings and assemble them
correctly or replace them
Check condition and clean/repair sealing faces
Replace gasket under top cover or safety relief
valve and/or sealing rings under plugs
Note! If symptoms persist, the pump must be taken out of service immediately. Contact your distributor.
3.20.1 Instructions for re-using and disposal
3.20.1.1 Re-use
Re-use or putting the pump out of service should only be undertaken after complete draining and
cleaning of the internal parts.
Note! When doing so, observe adequate safety regulations and take environmental
protection measures.
Liquids should be drained and following local safety regualtions the correct personal equipment
should be used.
3.20.1.2 Disposal
Disposal of the pump should only be done after it has been completely drained. Proceed according
to local regulations.
Where applicable please disassemble the product and recycle the parts’ material.
3.21 Maintenance instructions
3.21.1 General
This chapter only describes operations that can be performed on-site for normal maintenance.
For maintenance and repair requiring a workshop contact your distributor.
• Insufcient,wrongand/orirregularmaintenancecanleadtomalfunctionsinthepump,high
repair costs and long-term inoperability. Therefore, you should carefully follow the guidelines
given in this chapter.
During maintenance operations on the pump due to inspections, preventive maintenance or
removal from the installation, always follow the prescribed procedures.
Non-compliance with these instructions or warnings may be dangerous for the user and/or
seriously damage the pump/pump unit.
• Maintenanceoperationsshouldbeperformedbyqualiedpersonnelonly.Alwayswearthe
required safety clothing, providing protection against high temperatures and harmful and/or
corrosive fluids. Make sure that the personnel read the entire instruction manual and, in
particular, indicate those sections concerning the work at hand.
• SPXProcessEquipmentisnotresponsibleforaccidentsanddamagecausedbynon-
compliance with the guidelines.
• Personnelwhocarryacardiacpacemakershouldnotbeallowedtoworkonapumpequipped
with magnetic coupling!. The magnetic field is sufficiently strong to affect the proper operation
of a pacemaker, so keep a safe distance of at least 3 m.
• Donotcomeclose(notlessthan1m)tothemagneticcouplingwithsubjectsequippidwith
magnetic data carrier such as cheque cards, computer disks, watches etc. to avoid damage
and/or loss of information.
3.21.2 Preparation
3.21.2.1 Surroundings (on site)
Because certain parts have very small tolerances and/or are vulnerable, a clean work environment
must be created during on site maintenance. Particularly remove metallic chips or any dirt which
could be attracted by the components of the magnetic coupling.