10. After you have finished dispensing, press the START/STOP key to stop the pump
motor. Then measure the dispensed fluid’s volume using your calibrated beaker.
11. When you have determined the volume of the dispensed fluid, use the UP and
DOWN ARROW keys to enter this volume on the display.
12. Once the display shows the correct volume of fluid dispensed, press the
CALIBRATE key to complete the calibration. The display flashes and shows,
“CALIBRATED!”, and the red LED next to the CALIBRATE key lights up.
13. Now you can select a mode (RPM, FLOW RATE, TIME or TOTAL VOLUME)
(refer to Chapter 6). Once started, the fluid will be pumped according to the
calibration.
If you replace your tubing, make sure to press the TUBE ID key and select a
different tube id size using the UP or DOWN ARROW keys.
You can independently calibrate only one pump at a time. If
you have two pumps attached to the motor (even if they are stacked), only
one of the pumps can be calibrated.
For optimum FLOW RATE or TOTAL VOLUME accuracy, you will need to
recalibrate the tubing as a result of:
Using a different tube size (inner diameter)
Using a different tubing material
Using a tube of a different age (calibration should be checked
every 5-10 hours.)
Using a different back pressure in the tube
Changing the fluid being pumped
Changing the fluid’s temperature or density
Changing the RPM or FLOW RATE setting
Reversing the direction of pumping
Changing the clamp screw adjustment of the pump(s)
TIPS FOR OPTIMUM CALIBRATION ACCURACY:
Presented below are three useful techniques that will help you obtain optimum
calibration accuracy.
•Calibrate the unit at the speed you plan to use for normal pump operation.
When dispensing fluid during calibration, allow the pump to dispense the
largest total volume that your measurement equipment supports.
When measuring the dispensed fluid during calibration, calculate the
volume by measuring the fluid's mass with an accurate gram scale and
dividing this mass by the fluid's density.