Leica ASP300 Vacuum Cleaner User Manual


 
52
Instruction Manual V 1.3 – 12/2003
5.4 Reagent status
In the reagent status screen (Fig. 32) you can display and update the
actual reagent status of each station and you can set the stations
as full or empty.
Fig. 32
Warning thresholds that trigger
warning messages appear on a
red background.
To change a reagent status,
highlight the corresponding line
and press the corresponding
key symbol .
5. Operation
Reagent warnings
If the RMS is enabled, the instrument will give
you warning messages when reagents have
been used for too long.
Warning messages for normal reagents are
given based on up to 3 criteria; warning mes-
sages for paraffin wax on up to 6 criteria.
The warning messages are displayed at the
end of a cleaning cycle. From there, you can
go directly to the REAGENT STATUS screen.
In addition, warning messages (on a yellow
background) are displayed in the program
start screens.
Criteria for warning messages
For normal reagents, the following warning cri-
teria apply:
number of blocks processed
number of programs carried out
number of day lapsed
since the reagent has last been changed.
For paraffin wax, three additional criteria can
be applied:
number of blocks processed
number of programs carried out
number of days lapsed
since the wax has last been cleaned.