25
Leica ASP300
System diagnostics menu
If, during initialization, ‘Touch here
...’ is touched (refer Fig. 7) and sub-
sequently the supervisor pass-
word is entered, the SYSTEM
DIAGNOSTICS menu (Fig. 9) opens
up, giving access to standard in-
strument settings.
Attention!
Serious instrument malfunctions
may occur if any settings are
modified wrongly.
4. Setting up the instrument
Fig. 9
Fig. 10
The following functions can be selected:
1 - Access for service engineers only.
2 - Access INSTALLATION menu.
3 - Abort the current program.
4 - Abort the current program and delete the
current allocation of reagents to reagent
bottles and retort.
5 - Delete all programs and reset instrument
status (all lists will be empty).
6 - Store actual instrument status on floppy
disk.
7 - Display run log.
8 - Restart instrument.
To exit this menu, the instrument has to be re-
started - press: RESTART APPLICATION.
Press YES to confirm ‘ARE YOU SURE YOU
WANT TO ...’ (refer Fig. 10) .
• Initialization starts again from the screen
shown in Fig. 7.
3
4
2
1
5
8
7
6