Leica 106 Z Indoor Furnishings User Manual


 
Use the stray-light protection.
Danger to eyes from UV radiation.
During assembly work on xenon burners, always wear the
protective gloves and face protection provided.
Risk of explosion
Electrical components must be assembled at least 10 cm
away from the wall and from flammable substances.
UV radiationwarning
UV radiation can harm the eyes.
Never look into the light spot on the object level without a UV
protection screen.
Never look directly into the beam path (danger of blinding).
Always fill all filter positions with filter holders.
Do not select a white, strongly reflective background for the
object.
Never open the lamp housing when the lamp is switched on.
Danger of explosion, UV radiation, danger of blinding!
Avoid switching the lamp on and off frequently, as this
reduces its life.
Hot mercury burners light only after they have cooled down.
If possible, allow a new mercury burner to burn-in for 1–2
hours without interruption.
Install the mercury burner very carefully, and without
applying pressure.
Burners can burst very easily.
Never touch the glass parts of the mercury burner with your
hands. Fingerprints are burned on and reduce the light
quality. If necessary, carefully wipe off dust and fingerprints
using an alcohol-soaked cloth, then wipe dry.
1.4 Recommendations,
instructions
6
Leica lamp housing and mercury burner
1. Safety concept