Leica 106 Z Indoor Furnishings User Manual


 
To minimize the risk of explosion, replace a discolored
high-pressure mercury burner promptly when it has
reached the end of its rated life.
Follow the manufacturer’s specifications and the minute meter
on the supply unit!
̈Switch on the supply unit.
The safety starter ignites the burner. If the burner does not
ignite after several attempts, the burner could still be too hot
or may be burned out, and the safety starter then switches
off. Wait until the burner has cooled off or replace it with a
new one. New start:
̈Press the red button on the safety starter.
The safety starter becomes operational once again.
After switching on the supply unit, new mercury burners
must be adjusted immediately.
There is a danger of overheating for mercury burners
that are not adjusted.
Danger of explosion.
Replacing the safety starter
̈Turn the safety starter counterclockwise and remove it.
You can also use Osram safety starters for HBO 75W.
100W mercury burners must be powered by the ebq 100
supply unit.
Read the separate instruction manual and safety
instructions provided with this supply unit.
Leica lamp housing and mercury burner
4.2 ebq 100 supply unit
for 100W mercury
burners