SECTION 2—OPERATING THE BATH LIFT
1900/1900S Bath Lift 12 Part No. 1148113
Emergency Lowering
NOTE:Forthisprocedure,refertoFIGURE 2.6.
Intheev entofahandsetorcontrolboxfailure,theliftcanstillbeloweredbyusingone
ofthethreefollowingmethods,dependingonreasonthelifthasnotlowered:
Ifthebatteryisfine,butpendantisnotorcontrolboxhasfailedandtheliftneeds
tobelowered‐Pressthetipofaballpointpenintothesmallopeningbelowandtothe
rightofthestopbuttononthecontrolbox(Detail“A”).
Ifthebatteryhasfailedandtheliftneedstobelowered‐Pulldowntheredtag
locatedabovethebatterynearthepushhandleontheactuatorcolumn(Detail“B”).
NOTE:WeightorforceMUSTbeexertedupontheseatinorderforthisproceduretowork.
FIGURE 2.6 Emergency Lowering
Disinfecting the Lift
NOTE:Forthisprocedure,refertoFIGURE 2.7onpage 13.
ƽ WARNING
The seat and arms of the lift MUST be disinfected after each use with a bath.
ALWAYS wear protective gloves and eye protection when working concentrated
chemical disinfectants.
1. PrepareaworkingsolutionofInvacaredisinfectantbydoingthefollowing:
A. Mixconcentrateddisinfectantwithwatertoadilutionrationofhalfanounceof
disinfectantperonegallonofwater.
B. Transferthediluteddisinfectantintoalabeledspraybottle.
2. Positionthebathliftintothetub.RefertoSTEPS13to
16inTransferringinto/outof
theBathingUnitonpage 9.
3. Spraybothsidesoftheseatsurface,thesafetystrap,theshoulderstrap,andthearm
restsofthelift.RefertoDetail“A ofFIGURE 2.7.
4. Gentlywipethepushhandles,scale,handpendant,andanyothersurfacesthatwere
handledduringthetransportandbathingoftheresidentwithawashclothcontaining
thediluteddisinfectant.RefertoDetail“B”ofFIGURE 2.7.
5. Letthedisinfectantsitfortenminutes.
Press Ballpoint Pen Tip Here
for Emergency Lowering
DETAIL “A”
Control Box
DETAIL “B”
Pull Down Red Tag
for Emergency
Lowering
Actuator
Column