GE Analytical Instruments, Inc. ©2005H Page 4 of 16 DIN 30001 Rev. E
TOC,FLOW SWITCH UPGRADE 2.12CBI
TC ZERO CORRECTION
-1.3250000
ENTER
CLEAR
ENT = SAVE CHAR/NUM
CLR = RESTART/EXIT
A
RROWS = CHANGE CHA
R
FIGURE THREE: TOC ZERO OFFSET Screen
5.2.3. Enter 911 by using the arrows, and then press Enter.
ENTERING 911 WILL IN, NO WAY AFFECT THE VALUE OF THE
TC ZERO CORRECTION.
FACTORY ACCESS V. 0.00
ERASE ERR
S
SET SYRNGE
ENTER
CLEAR
PRNT SETU
P
SET EEPROM
SYRNGE MOV
SET SERVO
DUMP RAM
SET 4-20MA
SYRNG PUL
S
FIGURE FOUR: FACTORY ACCESS Menu
5.2.4. Using the arrows, select SETEEPROM, and press Enter.
5.2.5. Constants will be displayed in the sequence listed on the Constants Record
Form. Record each constant value on the form contained within this
document.
IMPORTANT:
It is critical that this Printout or Constants Record Form is retained for
comparison to the Constant Printout or Record Form at end of upgrade
procedure. These Constants will have to be checked and any variations
must be corrected after the upgrade is completed.
5.2.7. Record Dates and Percents for consumables on the Constants Record
Form. These will need to be reentered after upgrade.
5.2.8. From the MAIN MENU select CONSUMABLES, press Enter.
LAM
P
OXIDIZE
R
ENTER
CLEAR
A
CID
REAGENT FLUS
H
PUMP TUBING
FIGURE FIVE: Consumables Menu