GE 800 Water Pump User Manual


 
GE Analytical Instruments, Inc. ©2005H Page 11 of 16 DIN 30001 Rev. E
TOC,FLOW SWITCH UPGRADE 2.12CBI
Constants Record
CONSTANT VALUE
NAME
CONSTANT VALUE AFTER
UPGRADE
CONSTANT VALUE PRIOR
TO UPGRADE
C1 CONST:
Same as prior value
C2 CONST:
Same as prior value
C1 THERM A:
Same as prior value
C2 THERM.A:
Same as prior value
C1 THERM B:
Same as prior value
C2 THERM B:
Same as prior value
C1 THERM C:
Same as prior value
C2 THERM C:
Same as prior value
C1 XFER (60Hz):
Same as prior value
C2 XFER (60Hz):
Same as prior value
C1 COND:
Same as prior value
C2 COND:
Same as prior value
C1 TEMP:
Same as prior value
C2 TEMP:
Same as prior value
STEPPER 1:
Analyzer will automatically enter
this value – do not change.
STEPPER 2:
Analyzer will automatically enter
this value – do not change.
SOAK 1: 300
SOAK 2: 300
FLUSH 1: 60
FLUSH 2: 60
SYRINGE 1: 0.01445
SYRINGE 2: 0.01445
BAUD: 9600
STEPFILL 1: 20000
STEPFILL 2: 20000
STEPZERO: 24
DELAY: 24
CELL1 ZERO: 10
CELL2 ZERO: 10
COND SKIP: 8
A/D NOISE LEVEL: 60
TOC MODE:
Same as prior value
PRINTER COLUMS:
Same as prior value
PRINTER BREAK: 0
PRINTER TIME: 0
SERVO FILL PULSE:
Analyzer will automatically enter
this value – do not change
SERVO DRAIN PULSE:
Analyzer will automatically enter
this value – do not change
SLOPE WINDOW: 60
ANALOG MIN:
Same as prior value