GE Analytical Instruments, Inc. ©2005H Page 10 of 16 DIN 30001 Rev. E
TOC,FLOW SWITCH UPGRADE 2.12CBI
to save digits. After resetting the clock, Warning 35 will occur until the power
has been cycled two times.
5.7.2. Set Flowrates – The Firmware allows selection of different flow rates for on-
line sampling and grab (container) sampling. The grab flow rates can be
viewed and changed from the MAIN – SETUP/REAGENTS menu and from the
GRAB SAMPLING – SETUP/REAGENTS menu. Make sure to view and change
the desired flow rates for the sample mode being used.
5.7.3. From the SETUP menu, choose REAGENTS. From the REAGENTS menu, choose
ONLINE FLOW RATES. Set both reagent flowrates to the desired level. Select
the sampling mode.
5.7.3. From the SETUP menu, choose SAMPLE MODE.
5.7.3.1. Select ON LINE SAMPLING or GRAB SAMPLING. If GRAB SAMPLING is
selected, START TOC will access the GRAB SAMPLING menu, where
additional setup options may be implemented.
5.7.4. Select Frequency, only if AC FREQ. SETTING” is set to 50 for the 810 and 820
analyzers.
5.7.5.1. From SETUP, select 50/60HZ.
5.7.5.2. Select the appropriate Frequency for the analyzer.
5.7.6. Reset Acid/Oxidizer/UV Lamp/Pump Tubing dates and percentages.
5.7.6.1. Select CONSUMABLES from the MAIN Menu, press Enter.
5.7.6.2. Select ACID, OXIDIZER, UV LAMP, or PUMP TUBING, press Enter.
5.7.6.3. For the oxidizer, select SET % LEFT and enter the percent recorded on
the printout or the Constants Record Form. Use the arrow keys to
scroll through digits and press the Enter key to set the digits. Select
SET INSTALL DATE, press Enter. Use the arrow keys to scroll through
digits and press the Enter key to set the digits.
5.7.6.4. For the ACID, LAMP and PUMP TUBING, select SET % LEFT and enter the
percent recorded on the printout or the Constants Record Form. Use
the arrow keys to scroll through digits and press the Enter key to set
the digits.