5.0 CALIBRATION & MAINTENANCE
IMPORTANT: Before performing any calibration procedure,
ensure that fresh batteries are installed or use the optional
AC power adapter. Also, ensure that the analyzer is at room
temperature and will be sampling fresh air when powered
ON.
IMPORTANT: To main tain ac cu ra cy as listed in the Techni-
cal Characteristics Sec tion of this manual, the standards
used must be at least four times as accurate as the stated
accuracy of the Fyrite
®
INSIGHT.
5.1 B-Smart Sensors
The Fyrite
®
INSIGHT utilizes Bacharach’s new B-Smart Sensor tech-
nology for its CO sensor. The B-Smart sensor is marked with a 10-digit
calibration code that can be entered in the instrument's memory via the
keypad, or with Fyrite user software.
Benefits of B-Smart Sensors:
New sensors can be installed without applying gas for calibration.
Sensors can be pre-calibrated and installed when needed.
Sensors can be moved from one analyzer to another.
The B-Smart sensors should be calibrated by an authorized Bacharach
Service Center every 6 months to 1 year to assure that the analyzer contin-
ues to meet its published accuracy specifications. The B-Smart sensors,
however, can be calibrated in the field if your facility has the necessary
equipment and qualified personnel to perform the procedures described in
the following sections of this instruction manual.
5.2 Starting a Calibration
Start any calibration by performing the following:
1. With the analyzer power OFF, place the unit in fresh, ambient air;
then power on the analyzer.
After the installation,
• Open Fyrite
®
User Software via the desktop icon.
• Connect the instrument to the PC with a USB cable.
• Power up the unit.
• When first connected, a “Found New Hardware Wizard” screen will be
displayed.
Click on: "Next..."
"Next..."
"Continue anyway"
"Finish"
4.16 Fyrite
®
User Software Description
Fyrite
®
INSIGHT User Software
Set instrument time Sync instrument with computer's time
Calibration password Change 4 digit numeric code
Time meters Provides instrument & pump run time
B-Smart B-Smart calibration code entry
Saved measurements Downloads instrument memory to Excel
User/Customer ID User ID entry
Settings Change instrument set up parameter
Customer logo set up Load custom logos for print out. Limited to 384 x 192.
Custom fuel Bacharach can supply fuel codes for custom fuel. Consult
factory.
Update instrument software Remote upgrade of instrument software.