A&D MX-50 Thermostat User Manual


 
MS-70 / MX-50 / MF-50 / ML-50 53
12.
12.12.
12. Data Memory Function
Data Memory FunctionData Memory Function
Data Memory Function
The data memory function automatically stores each result when finishing a measurement.
MS-70 / MX-50 MF-50 ML-50
Maximum number 100 data 50 data 30 data
The stored data can be output to a printer at one time and can be output to a computer
using RsCom and Rskey that are Windows applications stored in the CD-ROM of
WinCT-Moisture or WinCT at one time.
The stored data can be deleted at one time.
The function can select either storing each result or not at data of the function table.
Data is stored at each measurement .......... data 1
Data is not stored........................................ data 0
When using the data memory function, MEM is displayed.
When displaying fUll , the function can not store the next data. The function can
store new data after deleting the stored data.
Caution
When pressing the STOP key during a measurement except manual mode,
the result is not stored.
Set data 1 before measurement, if it is necessary to store each result with data
memory function.
12.1.1.
12.1.1.12.1.1.
12.1.1. Preparation
PreparationPreparation
Preparation
This example selects "store result" at data of the function table.
1. Display the gram unit (of the weighing mode).
2. Press and hold the SELECT key to enter the
function table.
3. Press the SELECT key to display data .
3. Press the SELECT key several times and press
the or key to display data "1 .
5. Press the ENTER key to store the new settings.
Press the RESET key to return to the weighing mode.
MEM is displayed when the memory function is effective.
Memory function sign
Number of data
Full memory
0.000g
Cl adj
data "1
Press and hold
data
Store result
5-d
MEM
0
0.000g